Internal Standards Mixing Kit
CPI Home
CPI International Home
CPI International Privacy Policy
CPI International Site Map
About CPI
CPI International About Us
CPI International Company Info
CPI International Press
CPI International Management
CPI International Employment
Products
CPI International Product Index
CPI International Catalog
Promotions Promotions
CPI International Quick Order
CPI International Sale Items
Customer Service
CPI International Cust. Service
CPI International Tech Support
CPI International Sales
CPI International MSDS
CPI International News
CPI International Feedback
Contacts
CPI International Contact Us
CPI International Headquarters
CPI International Sales
CPI International Worldwide
Shop Online
CPI International Shop Online
CPI International Quick Order
MSDS
CPI International MSDS Info
Login
CPI International Log In

Shopping Cart




 

Internal Standards Mixing Kit

(instructions for use for TJA 4062-910)

The peristaltic pump tubing kit is designed for the automatic addition of a Scandium or Yttrium Internal Standard (I/S) and/or Li ionization suppressant solution to your samples.

1. Installation
To use the kit, attach the pump tubing to adjacent rollers on the peristaltic pump. Insert capillary tube from the ORANGE-GREEN-ORANGE winding into the bulk I/S or I/S + Li solution.  The capillary from the ORANGE-ORANGE-ORANGE winding is used to pump the samples.

2. Bulk Internal Standard Solution
This tubing kit will provide a sample dilution factor of about 20%.  For use with the ICAP 61E Trace Analyzer, or any of the axial instruments, it is recommended that the bulk solution be made up as follows:

For simple internal standard addition:                         5 ppm Sc or Y
For addition of I/S plus Li:
                                             5 ppm Sc or Y, plus 1000 ppm Li

The final concentration in the sample reaching the plasma will then be approximately 1ppm I/S (Sc or Y), and if present, 200ppm Li.

Note: The exact concentration of I/S and/or Li is not critical.  However, it is most important that sufficient bulk solution be available to complete the run.  If a new batch of bulk solution is made up, the instrument should be re-standardized prior to running any sample or check solutions.

3. Peristaltic Pump Speed
Due to the increased solution flow, the peristaltic pump speed should be reduced to 100 to 110 rpm from the usual 130 to 140 rpm.

Order:
4062-910 Internal Std. Mixing Kit, TJA 4pk On Sale in December - was $179, now $134

4062-914 Internal Std. Mixing Kit, 2-stop, org-wht, org-red, 4pk

4062-916 Internal Std. Mixing Kit,  2-stop, grn-grn, org-blk, 4pk

4062-947 Internal Std. Mixing Kit,  2-stop, 4pk

Please contact CPI Customer Service at 800-878-7654 with any questions regarding this part or its methods.