What Is an Internal Standard in ICP-MS?
An ICP-MS internal standard is an element or isotope added
at a consistent concentration to measured solutions. Its response is used to
monitor and correct changes in analyte signal caused by instrument drift,
sample-introduction variation, and matrix effects.
The internal standard should not be a target analyte or a
meaningful native component of the sample, and it should behave similarly to
the analytes it is used to correct.
Why Are Internal Standards Used in ICP-MS?
Internal standards help compensate for changes in instrument
response that could otherwise bias measured analyte concentrations.
- Instrument drift during an analytical sequence
- Changes in sample-introduction efficiency
- Nebulizer or spray-chamber variation
- Matrix-induced signal suppression or enhancement
- Physical differences between samples and
standards
- Short-term signal fluctuation
- Certain mass-dependent response changes
What Can Internal Standards Correct, and What Can’t They Correct?
|
Issue
|
Can Help?
|
Key Point
|
|
Instrument drift
|
Yes
|
Useful when the internal standard experiences similar
drift to the analyte.
|
|
Sample transport differences
|
Often
|
Can help compensate for nebulization and transport
variation.
|
|
Matrix suppression / enhancement
|
Often
|
Correction depends on the internal standard behaving
similarly to the analyte.
|
|
Polyatomic or isobaric interference
|
No / limited
|
Requires interference-specific correction or other
controls.
|
|
Incorrect standard preparation
|
No
|
Internal standardization cannot repair an incorrectly
prepared calibration standard.
|
|
Contamination
|
No
|
Use blanks, clean preparation, reagent control, and
appropriate laboratory practices.
|
How Are ICP-MS Internal Standards Selected?
Select an internal standard that behaves as much like the
analyte as practical while remaining distinguishable from the analyte and
absent from the sample at meaningful levels.
→ Method fit: Confirm the element is allowed
or appropriate for the analytical method and instrument configuration.
→ Sample absence: Avoid elements that are
naturally present at levels that materially change the added internal-standard
signal.
→ Mass / ionization behavior: Favor an
internal standard with sufficiently similar mass and/or ionization behavior to
track analyte response changes.
→ Interference risk: Confirm the selected
isotope is free from significant spectral or polyatomic interference and does
not create interference on target analytes.
→ Matrix compatibility: The internal standard
must remain soluble, stable, and chemically compatible with the calibration and
sample matrix.
→ Signal quality: Choose a concentration that
produces a stable, measurable signal without creating detector, contamination,
or interference problems.
What Are Common ICP-MS Internal Standard Elements?
Common internal-standard elements include scandium, yttrium,
rhodium, indium, terbium, holmium, lutetium, bismuth, lithium-6, gallium, and
germanium, depending on the method and instrument.
No single element is ideal for every analyte or matrix.
Multi-element ICP-MS methods often use several internal standards distributed
across the mass range so each analyte can be corrected using a suitable
reference element.
How Are Internal Standards Added?
Internal standards can be added directly to each blank,
calibration standard, QC material, and sample, or introduced online through a
separate pump channel and mixed with the sample stream before nebulization.
→ Direct addition: Requires consistent
volumetric addition and sufficient mixing to ensure homogeneity.
→ Online addition: Can improve consistency
during routine sequences, but tubing dimensions, uptake rate, mixing volume,
and stabilization time become important.
Whichever approach is used, the internal standard should be
added consistently to all relevant measured solutions, and the solution matrix
should be compatible with standards and samples.
What Concentration Should an ICP-MS Internal Standard Be?
There is no universal internal-standard concentration for
every ICP-MS method. The concentration should provide adequate precision while
avoiding contamination, detector issues, or interference problems.
For online addition, the stock solution may need to be more
concentrated than the effective concentration reaching the nebulizer because
the internal-standard pump channel may deliver a smaller flow than the sample
channel.
How Does Internal Standard Correction Work?
ICP-MS software monitors the internal-standard response
during analysis and compares it with the response observed during calibration.
If the internal-standard signal changes, the software applies a correction to
associated analytes according to the method and instrument settings.
The correction assumes that the analyte and assigned
internal standard are affected similarly. That assumption is why
internal-standard selection matters.
How Should Internal Standards Be Assigned to Analytes?
1. Start with the method — Confirm any
required internal-standard list or assignment rules.
2. Group analytes — Consider approximate mass
range and analytical behavior.
3. Review interferences — Check spectral,
polyatomic, and sample-matrix interference risks for each candidate.
4. Check sample presence — Avoid candidates
naturally present in the sample at meaningful levels.
5. Verify signal quality — Confirm the
internal-standard signal is stable and sufficiently intense.
6. Validate performance — Use QC recovery,
drift checks, matrix spikes, and representative samples.
7. Document assignments — Record the selected
internal standard for each analyte in the method or instrument settings.
What Can Changes in Internal-Standard Response Indicate?
A sudden or progressive change in internal-standard response
can be diagnostic information indicating a change in the measurement system.
- Nebulizer blockage or instability
- Pump tubing wear or uptake inconsistency
- Matrix deposition on cones
- Instrument drift
- Incorrect internal-standard addition
- Air bubbles or mixing problems
- High dissolved solids or matrix suppression
- Memory effects or carryover
- Interference on the internal-standard isotope
When Should a Laboratory Use a Custom Internal Standard Solution?
A custom internal standard solution can be useful when a
laboratory routinely runs the same ICP-MS method and needs a specific
combination of internal-standard elements, concentration, matrix, and volume.
→ Repeated workflows: The same
internal-standard mixture is used across routine analytical sequences.
→ Fewer preparation steps: A prepared mixture
can reduce repetitive manual additions and dilution steps.
→ Matrix requirements: The required matrix
differs from a catalog formulation.
→ Inventory planning: Routine volume needs are
predictable.
Custom formulations should still be evaluated for
compatibility, stability, and method suitability. The formulation should be
based on analytical requirements, not convenience alone.
What Information Is Needed to Specify a Custom ICP-MS Internal Standard
Solution?
→ Analytical method or application
→ Target analytes
→ Required internal-standard elements
→ Desired concentration of each internal
standard
→ Final matrix and acid concentration
→ Direct or online addition
→ Required volume and packaging
→ Regulatory or quality-system requirements
→ Instrument platform and sample matrix, when
relevant
CPI Support for ICP-MS Internal Standard Workflows
CPI International supports laboratories performing ICP-MS
and other elemental analysis workflows with:
For laboratories repeatedly preparing the same
internal-standard mixture, CPI custom inorganic standards allow the laboratory
to specify analytes, concentrations, matrix, volume, and other formulation
requirements.
Frequently Asked Questions
Does every ICP-MS method use the same internal standards?
No. Internal-standard choice depends on the method,
analytes, mass range, sample matrix, instrument configuration, and interference
profile.
Should internal standards be added to calibration standards and samples?
Yes, when internal standardization is used, they should be
added consistently to the relevant calibration standards, blanks, QC materials,
and samples according to the analytical method and laboratory procedure.
Can an analyte also be used as an internal standard?
Generally no. The internal standard should not be a target
analyte or be present in the sample at a level that materially changes the
added internal-standard signal.
Is an internal standard solution the same as a tuning solution?
No. A tuning solution evaluates instrument setup and
performance. An internal standard solution is used during analysis to monitor
and correct response changes.